Three phosphor plate systems and film compared for imaging root canals

Autor: A. C. Shearer, H.-G. Gröndahl, E. Mullane, Keith Horner, Tatiana V. Macfarlane
Rok vydání: 2001
Předmět:
Zdroj: International Endodontic Journal. 34:275-279
ISSN: 0143-2885
DOI: 10.1046/j.1365-2591.2001.00380.x
Popis: Aim The aim of this study was to compare three phosphor plate intraoral imaging systems (Digora (DA), Digident (DT), Denoptix (DX)) and E-speed conventional film (CF) for the imaging of root canals. Methodology Sixty extracted permanent teeth were exposed using CF, DA, DT and DX. The length of root canal visible from its most apical extent to a line drawn at the level of the cementoenamel junction was measured and this was expressed as a percentage of the distance between the radiographic apex and the cementoenamel line. All images were examined concurrently by two examiners. Films were viewed under standardized conditions and DA, DT and DX images were viewed directly from the monitor screen. The images on the monitor were enhanced to give the subjectively clearest image. Results The mean percentage of canal visible on CF was 90%, DA 78%, DT 81% and DX 83%. The difference was significant for CF–DA (P
Databáze: OpenAIRE