Calibration of isotope ratio mass spectrometry working standard for2H/1H ratio analysis

Autor: Shu Lim, Sara Bassilian, W. N. Paul Lee, E. Anne Bergner
Rok vydání: 1998
Předmět:
Zdroj: Journal of Mass Spectrometry. 33:627-630
ISSN: 1096-9888
1076-5174
DOI: 10.1002/(sici)1096-9888(199807)33:7<627::aid-jms677>3.0.co;2-7
Popis: A method is described for the calibration of an isotope ratio mass spectrometry working standard for 2H/1H analysis using highly purified (> 99.99% enriched) deuterium oxide (2H2O). Serial dilution of such a compound can be used to construct a standard curve for calibration purposes using the relationship [formula: see text] where APE is the atom percent excess. It was found that the determined rWR had a precision of +/- 1%. Most of the variation came from sample handling, which includes the serial dilution of the weighed deuterium oxide, the reduction of water to hydrogen gas and the introduction of the hydrogen gas into the mass spectrometer. The use of this calibration procedure allows the direct determination of the isotope ratio rSA, and APE from a standard curve over a wide range of enrichment. Unlike the reference standards V-SMOW and SLAP, the highly purified deuterium oxide is non-exhaustible and readily available. The highly purified deuterium oxide should be used in place of secondary standards to calibrate working references for 2H/1H analysis in biological studies.