Erratum to: ISCEV Standard for full-field clinical electroretinography (2015 update)

Autor: Michael Bach, Mitchell Brigell, Daphne L. McCulloch, Ruth Hamilton, Radouil Tzekov, Michael F. Marmor, Graham E. Holder
Rok vydání: 2015
Předmět:
Zdroj: Documenta Ophthalmologica. 131:81-83
ISSN: 1573-2622
0012-4486
DOI: 10.1007/s10633-015-9504-z
Popis: There is an error in Table 1 of the recently published ISCEV Standard for full-field clinical electroretinography (2015 update) [1]. Users of this standard should note that the filter settings for dark-adapted oscillatory potentials (OPs), as correctly specified in the text, require that the high-pass filter should remove frequencies of 75 Hz and below from the ERG waveform; the low-pass filter setting of 300 Hz or above is the same as for other ERG tests. Table 1 is provided below with the correct bandpass for OPs