MRI-based transfer function determination for the assessment of implant safety
Autor: | Janot P. Tokaya, J F Bakker, Peter R. Luijten, C.A.T. Van den Berg, Alexander J.E. Raaijmakers |
---|---|
Rok vydání: | 2017 |
Předmět: |
medicine.diagnostic_test
Computer science Magnetic resonance imaging Transfer function Pearson product-moment correlation coefficient 030218 nuclear medicine & medical imaging 03 medical and health sciences symbols.namesake 0302 clinical medicine Electric field Reciprocity (electromagnetism) Dielectric heating medicine symbols Radiology Nuclear Medicine and imaging Implant Image resolution 030217 neurology & neurosurgery Biomedical engineering |
Zdroj: | Magnetic Resonance in Medicine. 78:2449-2459 |
ISSN: | 0740-3194 |
DOI: | 10.1002/mrm.26613 |
Popis: | Purpose We introduce a new MR-based method to determine the transfer function (TF) for radiofrequency (RF) safety assessment of active implantable medical devices. Transfer functions are implant-specific measures that relate the incident tangential electric field on an (elongated) implant to a scattered electric field at its tip. The proposed method allows for TF determination with a high spatial resolution in relatively fast measurements without requiring dedicated bench setups from MRI images. Theory and Methods The principle of reciprocity is used in conjunction with the potential to measure currents with MRI to determine TF. Low-flip angle 3D dual gradient echo MRI data are acquired with an implant as transceive antenna, which requires minimal hardware adaptations. The implant-specific TF is determined from the acquired MRI data, with two different postprocessing methods for comparison. Results TFs of linear and helical implants can be determined accurately (with a Pearson correlation coefficient R ≥ 0.7 between measurements and simulations, and a difference in field at the tip ΔEtip ≤ 19%) from relatively quick (t |
Databáze: | OpenAIRE |
Externí odkaz: |