Autor: |
R. Botzenhardt, T. von Arnim, A. Matrai, Edzard Ernst, B. Weimer |
Rok vydání: |
1991 |
Předmět: |
|
Zdroj: |
Clinical Hemorheology and Microcirculation. 11:575-581 |
ISSN: |
1386-0291 |
DOI: |
10.3233/ch-1991-11603 |
Popis: |
Plasma viscosity and the viscous component of blood viscosity were measured in an oscillatory capillary viscometer (OCR-D) and compared to the blood viscosity determined in the Contraves LS-30 machine at identical nominal shear rates and to plasma viscosity readings obtained in the Coulter Harkness capillary viscometer. The OCR-D system yields lower reproducibility compared to the other methods. It also gives lower absolute viscositY readings for blood at low shear rates. Hemoconcentra tion, as induced in vivo by a standard ergometer exercise, is less strongly reflected in the low shear blood viscosity data from the OCR-D system compared to the analogue results obtained in the LS-30. These findings suggest, that the OCR-D method is less reproducible and not directly comparable with the "conventional" methods. Possibly the latter are best for routine while oscillatory methods may give a more detailed analysis of specific questions. |