Flow device to measure γ radioactivity and size of microparticles
Autor: | D. H. Frisch, Carolus M. Cobb, E. E. Frederick, A. T. Ramsey, Paul Bjorkholm, M. Annis |
---|---|
Rok vydání: | 1976 |
Předmět: | |
Zdroj: | Review of Scientific Instruments. 47:998-1004 |
ISSN: | 1089-7623 0034-6748 |
DOI: | 10.1063/1.1134811 |
Popis: | A description is given of the design and performance of an instrument which measures particle size and radioactivity of individual microparticles after uptake of a γ‐emitting radioactive tracer. A flow system allows the activated particles to be measured quickly and individually, giving statistically significant frequency distributions in a short time. The purpose of the flow device is to measure some property of particles of a given type, provided this property is correlated with the uptake of the tracer. The instrument sizes particles in the 5–150‐μm‐diam range, and detects γ rays of E≳30 keV. |