Physics and Astronomy (miscellaneous) Spectrometer business.industry Reference data (financial markets) General Engineering Physics::Optics General Physics and Astronomy Laser 01 natural sciences law.invention 010309 optics Laser linewidth Optics law 0103 physical sciences Femtosecond Calibration Spectral resolution 010306 general physics Spectroscopy business |
Zdroj: | Applied Physics B. 122 |
ISSN: | 1432-0649 0946-2171 |
DOI: | 10.1007/s00340-016-6473-1 |
Popis: | Two commercial femtosecond laser sources have been used to implement a dual-comb spectrometer tuneable across a spectral range from 1.5 to 2.2 μm. The optical linewidth of the comb modes was characterized for different time scales in order to estimate the achievable spectral resolution for an optimal acquisition time. The transmission spectra of three different gas samples were recorded, demonstrating good agreement with reference data. Frequency axis calibration was provided via the parallel monitoring of a reference sample. This technique allows an accurate calibration of the frequency axis of the spectrometer, with no need for stabilization or optical referencing of the frequency combs. Our set-up represents a good compromise for a compact and versatile dual-comb spectrometer based on commercially available parts with possible applications in trace-gas monitoring, remote sensing and spectroscopy of short-lived processes. |
Databáze: | OpenAIRE |
Externí odkaz: |
načítá se...