Self-centering of the Cook Celect inferior vena cava filter
Autor: | Daniel Howard Golwyn, Matthew Nathaniel Williams, John D. Regan, K. Dickey |
---|---|
Rok vydání: | 2021 |
Předmět: |
medicine.medical_specialty
business.industry medicine.medical_treatment Total knee arthroplasty Inferior vena cava filter Knee replacement General Medicine 030204 cardiovascular system & hematology medicine.disease Inferior vena cava Arthroplasty 030218 nuclear medicine & medical imaging Surgery Pulmonary embolism 03 medical and health sciences 0302 clinical medicine medicine.vein Filter (video) cardiovascular system medicine business Venous thromboembolism |
Zdroj: | American Journal of Interventional Radiology. 5:5 |
ISSN: | 2572-4606 2572-4614 |
DOI: | 10.25259/ajir_39_2020 |
Popis: | A patient who underwent placement of an inferior vena cava (IVC) filter before total knee arthroplasty with a history of venous thromboembolism following knee replacement 10 years prior. Difficulty releasing the filter during delivery resulted in approximately 17° of medial tilt of the filter apex. Follow-up computed tomography after arthroplasty 4 months after placement demonstrated that the filter self-centered in the IVC with subsequent straightforward retrieval. |
Databáze: | OpenAIRE |
Externí odkaz: |