Self-centering of the Cook Celect inferior vena cava filter

Autor: Daniel Howard Golwyn, Matthew Nathaniel Williams, John D. Regan, K. Dickey
Rok vydání: 2021
Předmět:
Zdroj: American Journal of Interventional Radiology. 5:5
ISSN: 2572-4606
2572-4614
DOI: 10.25259/ajir_39_2020
Popis: A patient who underwent placement of an inferior vena cava (IVC) filter before total knee arthroplasty with a history of venous thromboembolism following knee replacement 10 years prior. Difficulty releasing the filter during delivery resulted in approximately 17° of medial tilt of the filter apex. Follow-up computed tomography after arthroplasty 4 months after placement demonstrated that the filter self-centered in the IVC with subsequent straightforward retrieval.