KERMA-based radiation dose management system for real-time patient dose measurement

Autor: Kyung-Min Oh, Sung-Kwang Park, Sang-Sik Kang, Sang-Hee Nam, Ye-Ji Heo, Kyo-Tae Kim, Ji-Koon Park, Yongkeun Song
Rok vydání: 2016
Předmět:
Zdroj: Journal of the Korean Physical Society. 69:103-106
ISSN: 1976-8524
0374-4884
Popis: Because systems that reduce radiation exposure during diagnostic procedures must be developed, significant time and financial resources have been invested in constructing radiation dose management systems. In the present study, the characteristics of an existing ionization-based system were compared to those of a system based on the kinetic energy released per unit mass (KERMA). Furthermore, the feasibility of using the KERMA-based system for patient radiation dose management was verified. The ionization-based system corrected the effects resulting from radiation parameter perturbations in general radiography whereas the KERMA-based system did not. Because of this difference, the KERMA-based radiation dose management system might overestimate the patient’s radiation dose due to changes in the radiation conditions. Therefore, if a correction factor describing the correlation between the systems is applied to resolve this issue, then a radiation dose management system can be developed that will enable real-time measurement of the patient’s radiation exposure and acquisition of diagnostic images.
Databáze: OpenAIRE