Design, construction and performance testing of a 1.5 T cryogen-free low-temperature superconductor whole-body MRI magnet
Autor: | Yulin Wang, Sucong Wang, Ping Liang, Vladimir Brusic, Jie Zeng, Bingjian Liu, Yuwei Ge, Jianjun Zheng, Jingfeng Zhang, Chengbo Wang |
---|---|
Rok vydání: | 2023 |
Předmět: | |
Zdroj: | Superconductor Science and Technology. 36:045002 |
ISSN: | 1361-6668 0953-2048 |
DOI: | 10.1088/1361-6668/acb467 |
Popis: | This work describes the design, construction and testing of a 1.5 T cryogen-free low-temperature superconductor niobium–titanium whole-body magnet with a bore diameter of 850 mm that is suitable for clinical magnetic resonance imaging (MRI) applications. The magnet is actively shielded and passively shimmed to achieve confinement within a 0.5 mT stray magnetic field at 2.5/4 m radial/axial positions and 12.1 ppm field inhomogeneity over a 45 mm diameter of spherical volume. The magnet is conductively refrigerated using a two-stage Gifford–McMahon cryocooler and can be maintained at a steady temperature below 5.7 K during MRI scanning. The MRI scanner assembled with the designed cryogen-free magnet demonstrated stable performance and provided MRI images with good quality and high contrast. |
Databáze: | OpenAIRE |