Development and In Vitro Characterization of a New Immunoassay of Cardiac Troponin T

Autor: Andrew Remppis, A Borgya, U Essig, Klaus Hallermayer, U Geuss, T Scheffold, Hugo A. Katus, S Looser
Rok vydání: 1992
Předmět:
Zdroj: Clinical Chemistry. 38:386-393
ISSN: 1530-8561
0009-9147
DOI: 10.1093/clinchem/38.3.386
Popis: We describe a new one-step enzyme immunoassay of troponin T that uses two specific monoclonal antibodies and streptavidin-coated tubes as the solid phase. The monoclonal antibodies were obtained by conventional hybridoma technology, with human troponin T as antigen. The identity of the cardiac troponin T antigen was confirmed by analysis of the amino acid composition and by partial sequence analysis. The specificity of the monoclonal antibodies for cardiac troponin T was proved by immunoblot analysis and displacement curves. The capture antibody is labeled with biotin. The second antibody is conjugated to horseradish peroxidase (EC 1.11.1.7). The assay [Enzymun-TestR system (Boehringer Mannheim GmbH)] is performed in only 90 min at room temperature; the measuring range for troponin T is 0.1 to 15 micrograms/L. The assay shows excellent between-run precision (CV = 3.3-4.9%).