Installation, commissioning, and performance evaluation of Bhabhatron-II TAW – An Indian-made telecobalt unit

Autor: Shashi Bhushan Sharma, Gautam Sarma, Bhaveshwar Yadav, Mithu Barthakur, Pranjal Goswami, Shachindra Goswami
Rok vydání: 2020
Předmět:
Zdroj: Journal of Radiation Medicine in the Tropics. 1:38
ISSN: 2667-2537
DOI: 10.4103/jrmt.jrmt_12_20
Popis: Aim: The aim of this study was to report commissioning and performance evaluation of Bhabhatron-II TAW (an Indian made machine). Materials and Methods: Commissioning comprises of following quality assurance (QA) (a) electrical QA, (b) mechanical QA, (c) radiation QA and (d) survey of installation. Electrical QA were done for various components of the unit for its functionality. Mechanical QA were done and the measured data were verified as per guidelines of Atomic Energy Regulatory Board (AERB), India and vendor's specified values. Cylindrical ionization chamber, electrometer, ready pack films, water phantom, thermometer and barometer were used during radiation QA and survey of installation procedures. Moreover, the data so generated were fed to treatment planning system (TPS) for dose calculation and also kept in proper tabular manner for clinical use. Results: Couch and collimator parameters were found to comply the tolerance criteria. Source strength of 172.2 roentgen per minute at 1 meter (RMM) was measured with an absolute dose rate of 225.76 cGy/min at dmax for 10cm x 10cm field size at normal treatment distance (NTD). Shutter timer error of 0.004 min was measured and incorporated in the clinical data. A maximum of 3.66 % variation was noted for 300 (15W × 20) physical wedge transmission factor between measured and vendor specified value. Radiation leakage through beam limiting device (BLD) and outside BLD were found to be within the limits prescribed by AERB. Conclusion: Commissioning is a prerequisite process after installation of a radiation therapy machine for medical purpose. Bhabhatron-II machine has several advantages over other telecobalt machine. Features like high source storage capacity of machine head, one pair of asymmetric jaw, closure of jaws to 0 cm x 0 cm during radiation emergency and presence of one motorized wedge filter makes it unique and yet cost effective telecobalt machine..
Databáze: OpenAIRE