Autor: |
Michitaka Maruyama, Isao Tanaka, Eiji Sannomaru, Chiharu Urano, Takahiro Kanai, Nobu-Hisa Kaneko, Jun Honjo, Yuji Yoshino |
Rok vydání: |
2020 |
Předmět: |
|
Zdroj: |
2020 Conference on Precision Electromagnetic Measurements (CPEM). |
DOI: |
10.1109/cpem49742.2020.9191792 |
Popis: |
The influence of the ambient humidity on the output voltage of the nominal 10 V terminal of a prototype Zener dc voltage standard was investigated. In this measurement, the main module of the Zener standard driven with an external power supply unit was placed in a thermohygrostat chamber, and the output voltage was measured using a Josephson voltage standard system. When the humidity in the chamber was changed in the range 30%rh - 80%rh, no corresponding effects in voltage outputs were observed. |
Databáze: |
OpenAIRE |