SU-F-R-30: Interscanner Variability of Radiomics Features in Computed Tomography (CT) Using a Standard ACR Phantom

Autor: K. Latifi, Mikalai M. Budzevich, Geoffrey Zhang, M Shafiq ul Hassan, Eduardo G. Moros, Robert J. Gillies, Dylan C. Hunt
Rok vydání: 2016
Předmět:
Zdroj: Medical Physics. 43:3379-3379
ISSN: 0094-2405
Popis: Purpose: A simple approach to investigate Interscanner variability of Radiomics features in computed tomography (CT) using a standard ACR phantom. Methods: The standard ACR phantom was scanned on CT scanners from three different manufacturers. Scanning parameters of 120 KVp, 200 mA were used while slice thickness of 3.0 mm on two scanners and 3.27 mm on third scanner was used. Three spherical regions of interest (ROI) from water, medium density and high density inserts were contoured. Ninety four Radiomics features were extracted using an in-house program. These features include shape (11), intensity (22), GLCM (26), GLZSM (11), RLM (11), and NGTDM (5) and 8 fractal dimensions features. To evaluate the Interscanner variability across three scanners, a coefficient of variation (COV) is calculated for each feature group. Each group is further classified according to the COV- by calculating the percentage of features in each of the following categories: COV less than 2%, between 2 and 10% and greater than 10%. Results: For all feature groups, similar trend was observed for three different inserts. Shape features were the most robust for all scanners as expected. 70% of the shape features had COV
Databáze: OpenAIRE