First results from the Lund NMP particle detector system
Autor: | Christer Nilsson, Per Kristiansson, Jan Pallon, Marie Wegdén, Klas Malmqvist, N. Arteaga-Marrero, Mikael Elfman, Pavel Golubev, E.J.C. Nilsson |
---|---|
Rok vydání: | 2009 |
Předmět: |
High rate
Nuclear and High Energy Physics Microprobe Materials science Silicon Physics::Instrumentation and Detectors business.industry Detector chemistry.chemical_element STRIPS Ring (chemistry) Particle detector Ion law.invention Nuclear physics Optics chemistry law High Energy Physics::Experiment business Instrumentation |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 267:2065-2068 |
ISSN: | 0168-583X |
DOI: | 10.1016/j.nimb.2009.03.030 |
Popis: | The design and first results from a Double Sided Silicon Strip Detector (DSSSD) recently installed at the Lund Nuclear Microprobe facility (NMP) are presented. The detector has 64 sector strips and 32 ring strips, which in combination give more than 2000 detector cells, each with characteristics comparable with a standard surface barrier detector (SBD). The detector has been tested both with radioactive sources and with different ion beams and energies. The most striking features are the high rate virtually pile-up free operation and also the possibility of detailed measurement of angular distributions. (C) 2009 Elsevier B.V. All rights reserved. |
Databáze: | OpenAIRE |