Reproducibility and calibration of MMC-based high-resolution gamma detectors

Autor: Loredana Gastaldo, Achim Fleischmann, C. Pies, Sebastian Kempf, D. Hengstler, C. R. Bates, Stephan Friedrich, Christian Enss
Rok vydání: 2016
Předmět:
Zdroj: Applied Physics Letters. 109:023513
ISSN: 1077-3118
0003-6951
DOI: 10.1063/1.4958699
Popis: We describe a prototype γ-ray detector based on a metallic magnetic calorimeter with an energy resolution of 46 eV at 60 keV and a reproducible response function that follows a simple second-order polynomial. The simple detector calibration allows adding high-resolution spectra from different pixels and different cool-downs without loss in energy resolution to determine γ-ray centroids with high accuracy. As an example of an application in nuclear safeguards enabled by such a γ-ray detector, we discuss the non-destructive assay of 242Pu in a mixed-isotope Pu sample.