Elimination of the effect of aliphatic carboxylic acids in the determination of calcium by atomic absorption spectroscopy
Autor: | K. Szivos, E. Juhai, F. Trischler |
---|---|
Rok vydání: | 1983 |
Předmět: | |
Zdroj: | Mikrochimica Acta. 79:389-394 |
ISSN: | 1436-5073 0026-3672 |
DOI: | 10.1007/bf01204821 |
Popis: | In atomic absorption calcium determination performed in acetylene-air flame the interfering effect of citric, tartaric and oxalic acids has been investigated. It was found that the interfering effect could be eliminated by means of halogen acids (HCl, HBr, HI). The relative standard deviation is 0.5% for calcium. |
Databáze: | OpenAIRE |
Externí odkaz: |