The flame ionization detector a simple electrometer for linear, logarithmic and integral responses

Autor: V.E. Maier, R.A. Dewar
Rok vydání: 1964
Předmět:
Zdroj: Journal of Chromatography A. 15:461-470
ISSN: 0021-9673
DOI: 10.1016/s0021-9673(01)82804-1
Popis: A detailed description of a versatile DC electrometer for use with the flame ionization detector is given. The circuit provides for baseline compensation, and for logarithmic and integral responses, obtained by relatively simple electrical devices, all in the input circuit, as well as wide-range linear presentation in the usual way.