Compensation of the calibration drift in the spectrometers used for biomedical applications
Autor: | L. M. Selivanov, D. K. Kostrin, V. A. Gerasimov, V. A. Simon, A. A. Uhov |
---|---|
Rok vydání: | 2019 |
Předmět: |
Diffraction
Materials science Spectrometer Physics::Instrumentation and Detectors business.industry Astrophysics::Instrumentation and Methods for Astrophysics Physics::Optics Photodetector Photodiode law.invention Compensation (engineering) Optics law Position (vector) Calibration business Diffraction grating |
Zdroj: | XIV RUSSIAN-GERMANY CONFERENCE ON BIOMEDICAL ENGINEERING (RGC-2019). |
ISSN: | 0094-243X |
Popis: | One of the commonly used types of optoelectronic devices in biomedical applications is a spectrometer based on a concave diffraction grating and CCD photodetector located on the Rowland circle. Over time, the calibration of the device due to mechanical changes in the position of the elements in the optical scheme deteriorates. The paper deals with the method of compensation of calibration of the spectrometer, which does not require disassembly of the device. It is proposed to use a two-site photodiode that controls the position of the zero order of diffraction, which allows to automatically calibrate the device. |