Vergleich der Monitor- und Alternatorbefundung von digitalen Thoraxaufnahmen mittels eines computergestützten Verfahrens
Autor: | Karin A. Herrmann, M. Schätzl, Ulrich Fink, C. Pistitsch, Fink Bk, P. Miller, D. Heckermann, W. Kenn, M. F. Reiser |
---|---|
Rok vydání: | 1998 |
Předmět: | |
Zdroj: | RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren. 169:38-44 |
ISSN: | 1438-9010 1438-9029 |
DOI: | 10.1055/s-2007-1015047 |
Popis: | PURPOSE To compare softcopy and hardcopy reading of chest radiographs a software tool was designed for creating lesions with clearly defined size, location and contrast. METHOD An ROC study was performed using a set of chest radiographs with 300 simulated small nodules and linear details displayed on film and on a 1 K monitor. Six observers participated in the study. ROC curves were generated on the basis of 7200 decisions. RESULTS There was no significant difference in observer performance overall (AUC = Area under curve). Monitor: 0.84 +/- 0.08 vs. AUC, Film: 0.84 +/- 0.08). Slight, but not statistical significant differences were found in the detection of low-contrast lung nodules and of linear lung details. CONCLUSION Soft-copy reading using a high quality 1 K CRT display provides the same accuracy for the detection of subtle computer-simulated lesions as hardcopy reading. A software tool could be used for assigning optimal characteristics of CRT- and hardcopy displays using clearly defined lesions as a gold standard. |