Antegrade and retrograde embolisation of valves ensuing Transcatheter aortic valve implantation (TAVI)
Autor: | Bhaskar Ranganathan, Manoras Mathew Chengalath, Jeevesh John Thomas, Grace Maria George, Jose Chacko Periappuram |
---|---|
Rok vydání: | 2019 |
Předmět: |
medicine.medical_specialty
Transcatheter aortic business.industry 030204 cardiovascular system & hematology medicine.disease Surgery 03 medical and health sciences Stenosis 0302 clinical medicine medicine.anatomical_structure 030228 respiratory system Ventricle medicine.artery Ascending aorta medicine Device migration business Complication |
Zdroj: | IHJ Cardiovascular Case Reports (CVCR). 3:102-105 |
ISSN: | 2468-600X |
DOI: | 10.1016/j.ihjccr.2019.11.001 |
Popis: | There has been significant rise in Transcatheter aortic valve implantations (TAVI) worldwide. Displacement of the TAVI device is a potential life threatening complication. We present a case of 79 years old with severe aortic stenosis who underwent TAVI. When the valve got displaced into the ascending aorta, a second valve was deployed. The second valve gets displaced into the left ventricle. Failure of interventional efforts forced us to surgically explore to find extensive bruising and a dissected ascending aorta. Though the rate of device migration is quite low, a good understanding of the implicating factors will help the operator to avoid such a disaster. |