Effect Of The Band Width Of Receiving Transducer On Pulsed Doppler System

Autor: Hengyuan Zhao, Zezhang Hou, Yili Zhang, Qimin Chen
Rok vydání: 2005
Předmět:
Zdroj: [1990] Proceedings of the Twelfth Annual International Conference of the IEEE Engineering in Medicine and Biology Society.
DOI: 10.1109/iembs.1990.691113
Popis: In present study, the effect of the band width of receiving transducer on pulsed Doppler system was studied by usinp; an experimental system. The results of the study indicated that the stagger-tuned wide-band (1.4 MHz,-3dB) transducer(1) can give much more abundant information about blood flow than the narrow band (0.6 MHz, -3dB) transducer. The adequate hand-width of receiving transducer depends on the required signal/noise of transducer and the signal transmitting band.