Autor: |
Bernhard Laufer, Paul D. Docherty, Nour Aldeen Jalal, Sabine Krueger-Ziolek, Fabian Hoeflinger, Leonhard Reindl, Knut Moeller |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Current Directions in Biomedical Engineering. 8:616-619 |
ISSN: |
2364-5504 |
DOI: |
10.1515/cdbme-2022-1157 |
Popis: |
Tiffeneau manoeuvres are an important method in pulmonary function testing of the human lungs and can help to diagnose respiratory obstructions. Data from a motion capture system and a spirometer was used to evaluate Tiffeneau indexes which can theoretically be derived via a Smart-Shirt that incorporates three circumference measurements of the upper body. The mean error was 4.5% regarding the Tiffeneau indexes gained by the spirometer, indicating that clinical diagnosis of obstruction is potentially possible using a Smart-Shirt. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|