Katheterbruch eines implantierbaren zentralvenösen Infusionssystems mit Embolisation in die Arteria pulmonalis
Autor: | Meusers P, Kock Hj, Doetsch N, Krause U |
---|---|
Rok vydání: | 2008 |
Předmět: |
musculoskeletal diseases
medicine.medical_specialty business.industry medicine.medical_treatment Femoral vein General Medicine Catheter medicine.anatomical_structure Port (medical) Clavicle medicine.artery Pulmonary artery Medicine Embolization Radiology business Subclavian vein Central venous catheter |
Zdroj: | DMW - Deutsche Medizinische Wochenschrift. 114:1882-1884 |
ISSN: | 1439-4413 0012-0472 |
DOI: | 10.1055/s-2008-1066843 |
Popis: | A central venous catheter with a subcutaneous infusion port was implanted for chemotherapy into a 64-year-old woman with gastric lymphoma. Six weeks later, two weeks after a trivial trauma with hyperextension of the shoulder joint, it was found that the catheter had broken and its tip portion had embolized into the pulmonary artery: it was retrieved without difficulty via the femoral vein. This event suggests that pressure between clavicle and first rib can cause material fatigue in a silicon catheter. Such catheters should therefore not be implanted via the subclavian vein. |