Popis: |
Background:Hemostasis plays a crucial role during every surgery allowing for a bloodless operating field. Fast and effective surgery leads to a reduced risk of postoperative complications. One of the latest method for achieving homeostasis is the use of natural polysaccharide-based hemostatic powders. The study aimed to evaluate the biocompatibility according to the ISO 10993 standards of 4SEAL® Hemostatic powder.Methods:Chemical characterization (Headspace GC-MS, GC-MS, and ICP-MS), cytotoxicity, genotoxicity (MLA and AMES), endotoxin contamination, sensitization potential, intracutaneous reactivity, acute and subacute systemic toxicity with implantation, and pyrogenicity were evaluated to investigate the biocompatibility of the 4SEAL® Hemostatic powder. Studies were conducted according to ISO 10993 standards.Results:The biocompatibility requirements with accordance to ISO 10993-1 for 4SEAL® Hemostatic powder were met.Conclusions:4SEAL® Hemostatic powder is a promising new hemostatic agent with a wide range of potential applications and excellent biocompatibility. |