Dislocation rates of nasal versus oral airway exchange catheters for postoperative airway management. A prospective case series of 200 patients

Autor: L. Theiler, FM Roten, Fischer H, Wipfli M, Kleine-Brueggeney M, Greif R, Andreas Arnold, Steffen R
Rok vydání: 2019
Předmět:
DOI: 10.21203/rs.2.237/v1
Popis: BACKGROUND: The dislocation rate of oral versus nasal airway exchange catheters (AEC) in the postoperative care unit (PACU) are unknown. Our aim was to establish dislocation rates and to assess the usefulness of waveform capnography to detect dislocation. METHODS: In this non-randomized, prospective observational trial at the University Hospital Bern, Switzerland, we included 200 patients admitted to PACU after extubation via AEC, having provided written informed consent. The study was approved by the local ethical committee. AEC position was assessed by nasal fiberoptic endoscopy at beginning of PACU stay and before removal of the AEC. Capnography was continuously recorded via the AEC. Additional measurements included retching and coughing of the patient, and re-intubation, if necessary. RESULTS: Data from 182 patients could be evaluated regarding dislocation. Overall dislocation rate was not different between oral and nasal catheters (7.2% vs. 2.7%, p=0.16). Retching was more often noted in oral catheters (26% vs. 8%, pCONCLUSIONS: We found no difference in dislocation rate between nasal and oral position of an airway exchange catheter. However, nasal catheters seemed to be tolerated better. In the future, catheters like the staged extubation catheter may further increase tolerance. TRIAL REGISTRATION: The study was registered in a clinical study registry (ISRCTN 96726807) on 10/06/2010. KEYWORDS: Airway, extubation, intubation, airway exchange catheter, oral, nasal, postoperative, dislocation.
Databáze: OpenAIRE