Autor: |
Hiroyuki Nojiri, Emil Trakhtenberg, Yasuo Narumi, Dana Capatina, Zahirul Islam |
Rok vydání: |
2010 |
Předmět: |
|
Zdroj: |
Diamond Light Source Proceedings. 1 |
ISSN: |
2044-8201 |
Popis: |
The engineering of a dual-cryostat for a pulsed-magnet instrument at the Advanced Photon Source is presented. The dual-cryostat independently cools the magnet coil (using liquid-nitrogen) and the sample (using a closed-cycle refrigerator). Liquid-nitrogen cooling may allow a repetition rate of a few minutes for peak fields near 30 T. The system is unique in that the liquid-nitrogen cryostat incorporates a double-funnel vacuum tube passing through the solenoid's bore in order to preserve the entire angular range allowed by the magnet bore for scattering studies. Second, the use of a separate refrigerator for the sample allows precise positioning of samples in the bore while minimizing magnet vibrations propagating to the sample during pulsed-field generation. |
Databáze: |
OpenAIRE |