Successful retrieval of an entrapped coronary stent with its delivery system by small balloon dilatation at the entrapped site: A case report
Autor: | Hidenori Matsusaka, Toshiaki Ashihara, Toshihiko Kubo, Koshin Horimoto, Takashi Mizuki |
---|---|
Rok vydání: | 2015 |
Předmět: |
medicine.medical_specialty
business.industry Unstable angina medicine.medical_treatment technology industry and agriculture Percutaneous coronary intervention Stent equipment and supplies medicine.disease Balloon Surgery surgical procedures operative Restenosis Right coronary artery medicine.artery Conventional PCI Coronary stent medicine cardiovascular diseases Radiology business |
Zdroj: | Journal of Indian College of Cardiology. 5:67-70 |
ISSN: | 1561-8811 |
Popis: | A 73-year-old woman who had undergone percutaneous coronary intervention (PCI) with a drug-eluting stent in the proximal tortuous right coronary artery (RCA) 5 years previously was referred to our hospital because of unstable angina due to in-stent restenosis. During PCI, a drug-eluting stent with its delivery system became entrapped at the proximal edge of the previously placed stent, with its proximal half expanded into the aorta. The entrapped coronary stent was locked in the proximal tortuous RCA. We successfully retrieved the entrapped coronary stent with its delivery system by dilating a low-profile balloon at the entrapped site and dislocating the locking point (102 words). |