A device for controlling the energy of ionizing electrons in the low-energy range designed to form a complete mass spectrum of negative ions using a quadrupole mass spectrometer

Autor: A. G. Terentyev, A. V. Maltsev, R. V. Khatymov
Rok vydání: 2020
Předmět:
Zdroj: Industrial laboratory. Diagnostics of materials. 86:12-20
ISSN: 2588-0187
1028-6861
Popis: Operation with positive ions formed from the impact of high energy electrons (usually 70 eV, which exceeds the ionization energy of the molecules) is a standard mode for mass spectrometric detectors of most gas chromatography-mass spectrometric complexes (GC/MS) in the basic configuration. At the same time, we have shown that the option of setting the energy of ionizing electrons (maintained in the design of some serial devices) within the pre-ionization region also allows one to obtain mass spectra of negative ions (NI) which, in turn, significantly expands the analytical potential of standard equipment. The formation of NI occurs in the low-energy range of 0 – 15 eV due to resonant capture of electrons by molecules (REC). In contrast to positive ions, the intensity of NI formation sharply (resonantly) depends on the electron energy and this dependence is characteristic of each chemical compound. Both the relative intensity of the mass-peaks and, in general, the ionic composition of the formed mass spectrum of NI significantly depend on the electron energy. The problem of choosing the optimal energy of ionizing electrons providing the same efficiency of mass-spectrometric determination of all components of complex mixtures of dissimilar compounds is also associated with the features of negative ion formation during chromatography-mass spectrometric analysis. To address the problem, we propose a technique providing generation of complete (in NI composition and intensities) mass spectra of NI through repeated variation of the energy of ionizing electrons in a given range of 0 – 10 eV. Technical implementation of the technique [1] was carried out at the Design Bureau «Chromatec «(Yoshkar Ola, Russia) in the form of a special electronic device, which was tested in pilot operation as part of the gas chromatograph complex with a quadrupole mass spectrometer «Chromatec». We describe the principle of operation of the device and present the results of tests.
Databáze: OpenAIRE