Direct Detection of Pu-242 with a Metallic Magnetic Calorimeter Gamma-Ray Detector

Autor: Christian Enss, C. Pies, Stephan Friedrich, Sebastian Kempf, D. Hengstler, Loredana Gastaldo, Achim Fleischmann, C. Bates
Rok vydání: 2015
Předmět:
Zdroj: Journal of Low Temperature Physics. 184:351-355
ISSN: 1573-7357
0022-2291
DOI: 10.1007/s10909-015-1348-0
Popis: Cryogenic high-resolution $$\gamma $$ -ray detectors can improve the accuracy of non-destructive assay (NDA) of nuclear materials in cases where conventional high-purity germanium detectors are limited by line overlap or by the Compton background. We have improved the performance of gamma detectors based on metallic magnetic calorimeters (MMCs) by separating the 0.5 $$\times $$ 2 $$\times $$ 0.25 mm $$^{3}$$ Au absorber from the Au:Er sensor with sixteen 30- $$\upmu $$ m-diameter Au posts. This ensures that the entire $$\gamma $$ -ray energy thermalizes in the absorber before heating the Au:Er sensor, and improves the energy resolution at 35 mK to as low as 90 eV FWHM at 60 keV. This energy resolution enables the direct detection of $$\gamma $$ -rays from Pu-242, an isotope that cannot be measured by traditional NDA and whose concentration is therefore inferred through correlations with other Pu isotopes. The Pu-242 concentration of 11.11 $$\pm $$ 0.42 % measured by NDA with MMCs agrees with mass spectrometry results and exceeds the accuracy of correlation measurements.