Reliability of Holter Oximetry for Home Sleep Apnea Testing
Autor: | Redmond Shouldice, Conor Heneghan, Jordan C. Stern |
---|---|
Rok vydání: | 2008 |
Předmět: | |
Zdroj: | Otolaryngology–Head and Neck Surgery. 139 |
ISSN: | 1097-6817 0194-5998 |
Popis: | ObjectiveTo test the reliability of the Holter Oximeter for home testing of obstructive sleep apnea. Previous reports have shown a 96% correlation with simultaneous polysomnography and Holter Oximetry in the sleep laboratory. This study was designed to measure reliability of data obtained at home, as well as to obtain information from patients regarding comfort of the device.MethodsA prospective study of 120 consecutive patients (ages 5 to 85) presenting to an otolaryngology practice during a 4-month period with complaints of snoring or sleep apnea symptoms. Device: The Holter Oximeter produces an apnea hypopnea index (AHI) based on an automated processing method of a continuous electrocardiogram and pulse oximeter. The reliability of the test was determined by the number of tests completed without interruption due to patient discomfort, electrode or device failure.ResultsThere was 97% data recovery from the home testing device. Data failure was due to faulty memory cards in the device or surface electrod... |
Databáze: | OpenAIRE |
Externí odkaz: |