SODIUM AND POTASSIUM SENSORS BASED ON SEPARATED EXTENDED GATE FIELD EFFECT TRANSISTOR
Autor: | Hsiang-Ching Lee, Hsiang-Yi Hsu, Jyh-Ling Lin, Yuan-Lung Chin, Chang-Yeh Wu, Tai-Ping Sun |
---|---|
Rok vydání: | 2009 |
Předmět: | |
Zdroj: | Biomedical Engineering: Applications, Basis and Communications. 21:441-444 |
ISSN: | 1793-7132 1016-2372 |
DOI: | 10.4015/s1016237209001593 |
Popis: | A double potassium and sodium ions sensor based on separated extended gate field effect transistor (SEGFET) and a new entrapment method are proposed in this study. Having high sensitivity, reproducibility, and a simple manufacturing process, the sensors are quickly able to supply information and detect human blood within an abnormal range. The sensing membranes were fabricated using a Poly(vinyl chloride) carboxylated (PVC-COOH) matrix; sodium ionophore (B12C4) or potassium ionophore (valinomycin). The sensitivity of ion-selective biosensors is 42.55 mV/decade for Na+ between 1 M and 10-7 M, 42.75 mV/decade for K+ between 1 M and 10-6 M. |
Databáze: | OpenAIRE |