Application of a long-trace profiler in situ at TPS beamlines

Autor: Shang-Wei Lin, Huang-Ming Tsai, Huang-Wen Fu, Chih-Yu Hua, Chao-Chih Chiu, Kai-Yang Kao, Ming-Ying Hsu, Chia-Feng Chang, Hok-Sum Fung, Gung-Chian Yin
Rok vydání: 2022
Předmět:
Zdroj: Journal of Physics: Conference Series. 2380:012077
ISSN: 1742-6596
1742-6588
DOI: 10.1088/1742-6596/2380/1/012077
Popis: To achieve an ultrahigh-resolution with soft x-ray beamlines at the Taiwan Photon Source (TPS), the slope error of a highly precise grating must be at the level of 0.1 μrad (RMS) under thermal loading with various curvatures. Some optics generally operate under a high power density from an undulator magnet on a beamline. The thermal load introduces a thermal bump in the optics profile and degrades the beamline performance, such as the energy resolution and beam size. Hence, active X-ray mirrors are emerging as primary solutions to meet the performance requirements of TPS. First, active X-ray optics were assembled in the laboratory and bent to the desired surface using an in-house long-trace profiler (LTP). When the optics were installed in the beamlines, LTP in situ (ISLTP) was used to fine-tune the mirrors under various operating conditions. This paper introduces the design and construction of ISLTP with a precision of 0.1 μrad (RMS) developed to measure the mirror profile in soft x-ray beamlines. The ISLTP can provide feedback perturbations owing to localised heat loading from the X-ray beam and thermal changes in the ambient environment, and its performance and in situ measurements are presented.
Databáze: OpenAIRE