Autor: |
Ken Muldrew, George A. Sandison, John C. Saliken, Bryan Donnelly, John C. Rewcastle, R Baissalov |
Rok vydání: |
2000 |
Předmět: |
|
Zdroj: |
Medical Imaging: Image Processing |
ISSN: |
0277-786X |
Popis: |
Image guidance of cryotherapy is usually performed using ultrasound or x-ray CT. Despite the ability of CT to display the full 3D structure of the iceball, including frozen and unfrozen regions, the quality of the images is compromised by the presence of high density streak artifacts. To suppress these artifacts we applied Temporal Digital Subtraction (TDS). This TDS method has the added advantage of improving the gray scale contrast between frozen and unfrozen tissue in the CT images. Two sets of CT images were taken of a phantom material, cryoprobes and a urethral warmer (UW) before and during the cryoprobe freeze cycle. The high density artifacts persisted in both image sets. TDS was performed on these two image sets using the corresponding mask image of unfrozen material and the same geometrical configuration of the cryoprobes and the UW. The resultant difference image had a significantly reduced content of the artifacts. This TDS can be used in x-ray CT assisted cryotherapy to significantly suppress or eliminate high density x-ray CT streak artifacts by digitally processing x-ray CT images. Applying TDS in cryotherapy will facilitate estimation of the amount and location of all frozen and unfrozen regions, potentially making cryotherapy safer and less operator dependent. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|