Improvement of the performance of single-column ion chromatography by computerized detector signal processing and correlation chromatography

Autor: Mihkel Kaljurand, E. Urbas, Ülo Haldna
Rok vydání: 1994
Předmět:
Zdroj: Journal of Chromatography A. 660:147-152
ISSN: 0021-9673
DOI: 10.1016/0021-9673(94)85108-5
Popis: The performance of medium-quality ion chromatographic equipment was improved by using pseudo-random sample intoduction correlation chromatography (CC). The importance of the digital preprocessing of the detector signal prior to decorrelation was demonstrated for realization of the multiplex advantage of the CC method. The detection limits obtained for single-column ion chromatography using conductivity were 0.1, 0.12, and 0.17 ppm for Cl − , NO 3 − and SO 4 2− ions, respectively. Using the same apparatus without CC the detection limits are in the range 1–3 ppm.
Databáze: OpenAIRE