High-resolution MRI encoding using radiofrequency phase gradients

Autor: Vyacheslav Volotovskyy, Scott B. King, Jonathan C. Sharp, Boguslaw Tomanek, Qunli Deng
Rok vydání: 2013
Předmět:
Zdroj: NMR in Biomedicine. 26:1602-1607
ISSN: 0952-3480
DOI: 10.1002/nbm.3023
Popis: Although MRI offers highly diagnostic medical imagery, patient access to this modality worldwide is very limited when compared with X-ray or ultrasound. One reason for this is the expense and complexity of the equipment used to generate the switched magnetic fields necessary for MRI encoding. These field gradients are also responsible for intense acoustic noise and have the potential to induce nerve stimulation. We present results with a new MRI encoding principle which operates entirely without the use of conventional B0 field gradients. This new approach – ‘Transmit Array Spatial Encoding’ (TRASE) – uses only the resonant radiofrequency (RF) field to produce Fourier spatial encoding equivalent to conventional MRI. k-space traversal (image encoding) is achieved by spin refocusing with phase gradient transmit fields in spin echo trains. A transmit coil array, driven by just a single transmitter channel, was constructed to produce four phase gradient fields, which allows the encoding of two orthogonal spatial axes. High-resolution two-dimensional-encoded in vivo MR images of hand and wrist were obtained at 0.2 T. TRASE exploits RF field phase gradients, and offers the possibility of very low-cost diagnostics and novel experiments exploiting unique capabilities, such as imaging without disturbance of the main B0 magnetic field. Lower field imaging (
Databáze: OpenAIRE