A micro-discharge photoionization detector for micro-gas chromatography
Autor: | Shree Narayanan, Gary W. Rice, Masoud Agah |
---|---|
Rok vydání: | 2013 |
Předmět: | |
Zdroj: | Microchimica Acta. 181:493-499 |
ISSN: | 1436-5073 0026-3672 |
DOI: | 10.1007/s00604-013-1146-9 |
Popis: | We report on a small (20 × 10 mm) micromachined device for the detection of gases in micro-gas chromatography (GC). It incorporates a micro-discharge across a 20-μm gap, and a remote electrode in the micro cavity that generates an electrical signal corresponding to the photo-ionization of gaseous analytes in a stream of carrier gas. Multi-component mixtures were detected and the results compared to those obtained with a flame ionization detector. The minimum detectable limit is 350 pg.μL−1 of n-octane in air when applying a 1.4 mW discharge. The combination of wet etching of glass (as used for microfluidic channels) with a lift-off process for detector electrodes by a robust batch process results in a universal, non-destructive, and sensitive microdetector for micro-GC. |
Databáze: | OpenAIRE |
Externí odkaz: |