Application of a novel customized guide template in percutaneous pedicle screw placement: an in vitro study

Autor: Jian-Hua Chen, Sheng-Hai Wang, Guo Zheng, Su-Hua Wu
Rok vydání: 2018
Předmět:
Zdroj: EPiC Series in Health Sciences.
ISSN: 2398-5305
DOI: 10.29007/qzxg
Popis: In open surgery for spine stabilization, the pedicle screws are often placed using patient-specific guide templates since they can improve the screw placement accuracy and simplify surgery. However, the conventional fit-and-lock template requires full exposure of the bony structures and is therefore not suitable for minimally invasive procedures. In this study, we constructed a novel guide template for percutaneous pedicle screw placement. Its feasibility and the accuracy of template-assisted pedicle screw placement were assessed using an excised calf spine model. As a result, all inserted guide wires and 27 of 30 simulated screws (90%) were completely in the pedicle. The remaining screws (10%) penetrated by less than 2 mm. The average distance and angular deviations of the guide wires were 1.46±0.60 mm and 1.10±0.84°. Our study demonstrate that this novel guide template is technically feasible and enhances the accuracy of percutaneous pedicle screw placement. Moreover, it may simplify the surgery and minimize intraoperative radiation. Further research on its clinical applications is warranted.
Databáze: OpenAIRE