Evaluation of a charge injection device array

Autor: Roger L. Easton, Zoran Ninkov, Chen Tang
Rok vydání: 1994
Předmět:
Zdroj: SPIE Proceedings.
ISSN: 0277-786X
Popis: A Charge Injection Device Technologies 512 X 512 CID array has been tested at liquid nitrogen temperature to assess its suitability for use in low light applications. The CID-38 chip used was installed in a SICam dewar/electronics from CIDTEC. For this particular system the noise floor was measured to be approximately equals 260 electrons rms for a single readout. Using the nondestructive readout option an approximately equals (root)N, where N is the number of readouts, improvement in this noise figure can be achieved. The average dark current at 77 degree(s)K was not measurable in a three-hour dark integration.© (1994) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.