Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the LNE–LNHB, France and the BIPM in60Co gamma radiation

Autor: M Donois, David T Burns, F. Delaunay, C Kessler
Rok vydání: 2013
Předmět:
Zdroj: Metrologia. 50:06019-06019
ISSN: 1681-7575
DOI: 10.1088/0026-1394/50/1a/06019
Popis: An indirect comparison has been made of the standards for absorbed dose to water in 60Co radiation of the Laboratoire National de Metrologie et d'Essais–Laboratoire National Henri Becquerel (LNE–LNHB), France and of the Bureau International des Poids et Mesures (BIPM). The measurements at the BIPM were carried out in April 2013. The comparison result, based on the calibration coefficients for two transfer standards and evaluated as a ratio of the LNE–LNHB and the BIPM standards for absorbed dose to water, is 0.9971 with a combined standard uncertainty of 3.9 × 10−3. The results are analysed and presented in terms of degrees of equivalence for entry in the BIPM key comparison database. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Databáze: OpenAIRE