A non-contact ECG sensing system with motion artifact equalization
Autor: | Wai-Chi Fang, Ding-Yu Huang, Shanqing Gao |
---|---|
Rok vydání: | 2016 |
Předmět: |
Engineering
Signal processing Artifact (error) business.industry 020208 electrical & electronic engineering 010401 analytical chemistry Equalization (audio) 02 engineering and technology 01 natural sciences Capacitance 0104 chemical sciences Compensation (engineering) Normal distribution Distortion 0202 electrical engineering electronic engineering information engineering Electronic engineering business Equivalent input |
Zdroj: | BioCAS |
DOI: | 10.1109/biocas.2016.7833777 |
Popis: | This paper presents an electrocardiograph (ECG) non-contact sensing system that compensates the high-frequency noises and the distortion due to motion artifact via two-channel interleaved electrodes in concert with proportional design and equalization algorithm. Specifically, the equivalent input capacitances of the sensing electrodes are deliberately designed to achieve a fast, low-cost, and real-time compensation signal processing. This proposed system has been implemented with a worst case of 2% unit capacitance mismatch and 25% motion artifact with a normal distribution, where the equalization can still achieve average 98.6% high accuracy. |
Databáze: | OpenAIRE |
Externí odkaz: |