EPR dosimetry for actual and suspected overexposures during radiotherapy treatments in Poland

Autor: Andrzej Wojcik, I. Clairand, J. Sadlo, J. Rostkowska, W. Stachowicz, A. Mazal, J. Sluszniak, S. Gozdz, J. Michalik, François Trompier, Wojciech Bulski, A. Kulakowski
Rok vydání: 2007
Předmět:
Zdroj: Radiation Measurements. 42:1025-1028
ISSN: 1350-4487
Popis: EPR dosimetry on bone samples was recently used for actual and suspected overexposures during radiotherapy treatments performed in Poland. In 2001 five breast-cancer patients undergoing radiotherapy in the Bialystok Oncology Center, Poland, were overexposed. The overexposure was due to a defective safety interlock and an obsolete safety system of the linear accelerator. For the three most exposed patients, pieces of rib bones removed during surgical reconstruction of the chest wall and skin transplantation allowed an estimation of the accident doses by electron paramagnetic resonance (EPR) spectrometry. The doses delivered during the accident were as high as 60–80 Gy. In 2005, a patient treated in Kielce Holy Cross Cancer Center exhibited similar deep necroses of the chest wall but 6 years following a “standard upper mantel fields” radiotherapy for Hodgkin's disease. In order to investigate the possible late effect of an overexposure as necrosis origin, the delivered dose was afterward estimated by EPR dosimetry performed on a rib sample.