Use of a DC SQUID receiver preamplifier in a low field MRI system

Autor: David J. Lurie, H.C. Seton, James M. S. Hutchison, D. M. Bussell
Rok vydání: 1995
Předmět:
Zdroj: IEEE Transactions on Appiled Superconductivity. 5:3218-3221
ISSN: 1051-8223
DOI: 10.1109/77.403276
Popis: We have used tuned receiver coils coupled to a dc SQUID preamplifier in a small scale magnetic resonance imaging (MRI) system operating at 425 kHz (B/sub 0/=0.01 T). The coil and SQUID are cooled to 4.2 K in a modified biomagnetic cryostat. The modifications provide transparency to rf signals originating outside the cryostat while maintaining an acceptably low liquid helium boiloff rate. By applying negative feedback we have damped low loss, high Q-factor SQUID input circuits to achieve a useful imaging bandwidth of over 2 kHz. >