Simultaneous determination of organotin, organolead, and organomercury compounds in environmental samples using capillary gas chromatography with atomic emission detection
Autor: | Yan Liu, Werner F. Beckert, Viorica Lopez-Avila, Marcela Alcaraz |
---|---|
Rok vydání: | 1994 |
Předmět: | |
Zdroj: | Journal of High Resolution Chromatography. 17:527-536 |
ISSN: | 1521-4168 0935-6304 |
DOI: | 10.1002/jhrc.1240170705 |
Popis: | As part of a continuing evaluation of new analytical and sample preparation techniques conducted by the US Environmental Protection Agency (EPA), the use of capillary gas chromatography with atomic emission detection (GC-AED) for the simultaneous determination of organotin, organolead, and organomercury compounds in environmental samples was investigated. Pentylmagnesium bromide was used to pentylate ionic organotin, organolead, and organomercury compounds; the pentyl derivatives were then separated by GC and determined by AED. Several important GC-AED parameters, including the type of injector inlet, carrier gas flow rate, and helium make-up gas flow rate, were optimized for the simultaneous determination of these organometallic compounds. Their minimum detectable concentrations were approximately 1.0 to 2.5 ng/mL using a 0.5-μL on-column injection. The calibration curves exhibited good linearity between 2.5 and 2500 ng/mL for organotin and organolead compounds, and between 2.5 and 10000 ng/mL for organomercury compounds. |
Databáze: | OpenAIRE |