Performance characteristics of a chemical imaging time-of-flight mass spectrometer

Autor: Nicholas Winograd, Robert M. Braun, John C. Vickerman, K. F. Willey, Steve J. Mullock, Paul Blenkinsopp, Clive Corlett
Rok vydání: 1998
Předmět:
Zdroj: Rapid Communications in Mass Spectrometry. 12:1246-1252
ISSN: 1097-0231
0951-4198
DOI: 10.1002/(sici)1097-0231(19980930)12:18<1246::aid-rcm316>3.0.co;2-c
Popis: A chemical imaging time-of-flight secondary ion mass spectrometer is described. It consists of a liquid metal ion gun, medium energy resolution reflectron mass analyzer, liquid nitrogen cooled sample stage, preparation chamber and dual stage entry port. Unique features include compatibility with laser postionization experiments, large field of view, cryogenic sample handling capability and high incident ion beam current. Instrument performance is illustrated by the characterization of scanning electron microscopy grids, silver and functionalized polystyrene beads and the postionization of an organic overlayer on a gold substrate.