Experimental Validation of a Numerical Model of an Axial Circulatory Assist Blood Pump
Autor: | E. P. Banin, F. D. Sorokin, S. V. Skoryukov, Alexandre Gouskov, A. E. Krupnin |
---|---|
Rok vydání: | 2019 |
Předmět: |
business.industry
Rotor (electric) Computer science Design of experiments 0206 medical engineering Flow (psychology) Biomedical Engineering Medicine (miscellaneous) Magnetic bearing 02 engineering and technology Mechanics Blood flow Computational fluid dynamics 020601 biomedical engineering law.invention Physics::Fluid Dynamics Blood pump Medical Laboratory Technology law Working fluid business |
Zdroj: | Biomedical Engineering. 53:77-81 |
ISSN: | 1573-8256 0006-3398 |
DOI: | 10.1007/s10527-019-09881-5 |
Popis: | This study continues our systematic exploration of the design of an axial circulatory assist blood pump with a rotor on magnetic bearings. The work reported here addresses experimental validation of a numerical model of flow in the flow part of the axial blood pump. The study focuses on design of the test rig, design of experiments, and modeling of blood flow in the flow part of the pump using computational fluid dynamics. The working fluid was distilled water. The flow-pressure characteristics obtained experimentally were compared with those from mathematical modeling. A satisfactory agreement between the results was obtained, such that the mathematical model can be used for future simulation of the operation of the pump and further optimization of the main components. |
Databáze: | OpenAIRE |
Externí odkaz: |