Nuclear and High Energy Physics business.industry Astrophysics::High Energy Astrophysical Phenomena Detector Implosion Cosmic ray Astrophysics::Cosmology and Extragalactic Astrophysics Scintillator Laser law.invention Nuclear physics Time of flight Optics law Neutron business Instrumentation Ultrashort pulse |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 491:474-477 |
ISSN: | 0168-9002 |
DOI: | 10.1016/s0168-9002(02)01176-2 |
Popis: | The time response functions of an ultrafast quenched plastic scintillation detector used to measure neutron time of flight spectra were calibrated by utilizing cosmic rays and implosion neutrons from DT-filled capsules at the Shenguang II laser facility. These sources could be regarded as δ function pulses due to their much narrower time widths than those of the time response functions of the detection system. The results showed that the detector responses to DT neutrons and to cosmic rays were 1.18 and 0.96 ns FWHM, respectively. |