Performance of the Accelerate Diagnostics PhenoTM system with resin-containing BacT/ALERT® Plus blood culture bottles
Autor: | Kevin C. Hazen, Danielle C. Robinson, Kimberly Starr |
---|---|
Rok vydání: | 2019 |
Předmět: |
0301 basic medicine
Microbiology (medical) Antiinfective agent medicine.medical_specialty medicine.diagnostic_test business.industry Bact alert 030106 microbiology Antimicrobial susceptibility General Medicine medicine.disease Patient care Rapid identification 03 medical and health sciences 0302 clinical medicine Infectious Diseases Blood culture bottles Bacteremia Internal medicine medicine Blood culture 030212 general & internal medicine business |
Zdroj: | Diagnostic Microbiology and Infectious Disease. 94:122-128 |
ISSN: | 0732-8893 |
DOI: | 10.1016/j.diagmicrobio.2019.01.002 |
Popis: | Bacteremia and septicemia require rapid identification (ID) and antimicrobial susceptibility testing (AST) to start targeted, appropriate therapy. To answer this need, Accelerate Diagnostics, Inc., developed the Accelerate Pheno™ system (AXDX), a fast ID and phenotypic AST platform. Performance of a pre-FDA clearance version of AXDX was evaluated using 261 positive BacT/ALERT® Plus bottles and compared with standard of care (SOC). Average time to ID was reduced by 24.9±6.9 h and AST by 36.7±18.9 h compared with SOC. AXDX reports ID and AST of blood pathogens in 1.9 and 7.1 h. Positive percent agreement and negative percent agreement of AXDX ID were 94.5% and 98.9%, respectively. AXDX AST had an essential agreement of 96.5% and categorical agreement of 94.6% with 4 major errors and 7 very major errors. AXDX performance was acceptable for all 3 bottle types. Rapid ID and AST with AXDX could impact patient care and antimicrobial stewardship. |
Databáze: | OpenAIRE |
Externí odkaz: |