Microfocus X-ray sources for 3D microtomography

Autor: Scott J. Wilderman, Michael J. Flynn, Sean M. Hames, David A. Reimann
Rok vydání: 1994
Předmět:
Zdroj: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 353:312-315
ISSN: 0168-9002
DOI: 10.1016/0168-9002(94)91664-0
Popis: An analytic model for the performance of cone beam microtomography is described. The maximum power of a microfocus X-ray source is assumed to be approximately proportional to the focal spot size. Radiation flux penetrating the specimen is predicted by a semi-empirical relation which is valid for X-ray energies less than 20 keV. Good signal to noise ratio is predicted for bone specimens of 0.1 to 10 mm when scanned at the optimal energy. A flux of about 1 × 10 10 photons/mm 2 /s is identified for 0.2 mm specimens. Cone beam volumetric microtomography is found to compare favorably with synchrotron based methods.