Anatomy for right ventricular lead implantation

Autor: Carsten W. Israel, Sona Tribunyan, S. Yen Ho, José A. Cabrera
Rok vydání: 2022
Předmět:
Zdroj: Herzschrittmachertherapie + Elektrophysiologie. 33:319-326
ISSN: 1435-1544
0938-7412
DOI: 10.1007/s00399-022-00872-w
Popis: To understand the position of a pacing lead in the right ventricle and to correctly interpret fluoroscopy and intracardiac signals, good anatomical knowledge is required. The right ventricle can be separated into an inlet, an outlet, and an apical compartment. The inlet and outlet are separated by the septomarginal trabeculae, while the apex is situated below the moderator band. A lead position in the right ventricular apex is less desirable, last but not least due to the thin myocardial wall. Many leads supposed to be implanted in the apex are in fact fixed rather within the trabeculae in the inlet, which are sometimes difficult to pass. In the right ventricular outflow tract (RVOT), the free wall is easier to reach than the septal due to the fact that the RVOT wraps around the septum. A mid-septal position close to the moderator band is relatively simple to achieve and due to the vicinity of the right bundle branch may produce a narrower paced QRS complex. Special and detailed knowledge is necessary for His bundle and left bundle branch pacing.